Firebreak Valve
Features
Bi-directional design for versatile application.
Compact size makes it discreet within oxygen supply tubing.
Specifically engineered to help prevent fires in oxygen tubing.
Features a thermal fuse that automatically shuts off gas flow if downstream tubing catches fire.
Enhances patient safety during oxygen therapy.
Compatible with all oxygen sources: tanks, concentrators, and wall-mounted systems.
Designed for use with a single oxygen source.
Note: Does not restrict oxygen flow unless heat or fire is detected.
Applications:
| Firebreak Valve | |||
| Resistance to Flow | 1 kPa @ 2 l/min | ||
| 2 kPa @ 5 l/min | |||
| 16 kPa @ 15 l/min | |||
| Flow Rate | 0.3 to 25 l/min | ||
| Max Operating Pressure | 400 kPa | ||
| Max Internal Leakage Following Activation | No leakage | ||
| Max External Leakage Following Activation | No leakage | ||
| Operating Temperature | 0 to 50 °C (32 to 122 °F) | ||
| Transit & Storage Temperature Range | -20 to 60 °C (-4 to 140 °F) | ||
| Gas Type | Oxygen / oxygen-enriched air | ||
Applications:
The firebreak valve ensures safety by preventing oxygen leaks and protecting against tubing fires. It is compatible with all types of oxygen sources and is specifically designed to safeguard patients during oxygen therapy.